Reliability of results, ease of use and immediate information.

Complete range of tests for the control of autoclave sterilization. Bowie & Dick Test, Helix Test and chemical indicators, check the correct penetration of the steam inside the sterilization load, thus ensuring an adequate operating performance.
Image

Steam sterilization integrator control

Image
Verify the effectiveness of the sterilization process, analysing the critical variables such as time, temperature and presence of steam.

Bowie & Dick Test

Image

Verifies the complete steam penetration in the porous loads such as gauze, bandages, gowns and fabrics.

Helix test

Image
Ascertains the correct degree of steam penetration of the bodies in close lumen, such as turbines and contra-angles.
BOWIE&DICK
FUNCTION FREQUENCY CYCLE TYPE
Check the correct steam penetration in the porous bodies Daily
At the beginning of each shift
In the event of displacement of the autoclave
After any repair or maintenance intervention
B type cycle at 134°C for 3.5 minutes
HELIX TEST
FUNCTION FREQUENCY CYCLE TYPE
Check the correct steam penetration in the close lumen bodies Giornaliera
Daily
At the beginning of each shift
In the event of displacement of the autoclave
After any repair or maintenance intervention
B type cycle at 134°C for 3.5 minutes
B type cycle at 121°C for 15 minutes
CHEMICAL INDICATORS
FUNCTION FREQUENCY CYCLE TYPE
Verify the effectiveness of the sterilization process Single indicator for packaging for each sterilization load B type cycle at 134°C for 3.5 minutes
B type cycle at 121°C for 15 minutes